Agilent 33220A DC offset + OPAMP offset is not as it should be when i use labview

$
0
0

Hi there,
The output of Agilent 33220A can be maximum +5V. I need more than +5V, so i made a simple DC amlifier circuit connected to the 50ohms output of the instrument. (I need 2*Vin)

When i set manually the output of the intrument as "High Z" then all is fine.
Then amplification works as desired.: I tried 0...4V input->0...8V output.

 

Then I tried to set output in the Labview as follows:
Offset voltage = 1V.


DC amplifier board disconnected:
When I connect my voltmeter to the output of the function generator i measure the same as I configured in labview: 1V

DC amplifier board connected:
When I connect my amplifier circuit to the 50 ohms output of functioin generator,  then i get some strange amplification value. 2.59V
Measurement voltage should be 2V.

I tested amplifier circuit and it works fine without labview.
In short I can't get higher than 5V in the output of amplifier circuit no matter what i do in labview, but when i set things manually all is fine.

 

Have you got any ideas why this happens?
agilent.jpg