Agilent 33500B - "Too much data" when loading

$
0
0

Hi,

 

I am controlling an Agilent 33522B using VISA and am having some trouble trying to load a 1MSa arbitrary waveform. When I try and load any more than 70KSa I get the following errors:

" -223: Too much data
-101: Invalid character...
...Repeat -101: Invlaid character x 15...
-350: Queue overflow"

I am loading the data down in Definite length binary block format. 

A second bug I am seeing is that although I am loading 70KSa (140KB) to the instrument, it displays the arbitrary waveform as being 140KSa long. Is there some setup element where I have to configure that each sample is two bytes long? From reading Agilent's documentation I believed that the device can accept samples only in I16 or Double forms?

If you have any solutions or similar experiences please let me know as soon as possible.

Thanks,
Darragh