Quantcast
Channel: Instrument Control (GPIB, Serial, VISA, IVI) topics
Viewing all articles
Browse latest Browse all 5566

A weird communication failure occurred using a Keysight oscilloscope

$
0
0

I use Keysight oscilloscopes, models are DSOX3024T and DSOX4024 respectively,Labview and Visa are 2017 versions.Running the waveform reading program 20,000 times in a row on one computer is fine, but running it about 50 times on other computers will cause the oscilloscope to freeze,VISA error 0xBFFF003E(DSOX4024 about 200 times).The study found no problems with computers without USB3.0.

Because it is difficult to return the computer to USB2.0 driver, SO I used LAN communication, DSXO4024 before running the program can execute 2000 times or so crash, DSOX3024T is in the experiment. Share this experience with everyone, have similar experience can discuss


Viewing all articles
Browse latest Browse all 5566

Trending Articles



<script src="https://jsc.adskeeper.com/r/s/rssing.com.1596347.js" async> </script>