GPIB connection with Keysight B1500A analyzer - problems with Agilent 82357B controller

$
0
0

Hi NI community,

I am trying to control our Keysight B1500A analyzer remotely via GPIB using the LabView Development Environment.

 

I followed the suggested way of establishing a connection with the instrument using an Agilent 82357B USB-GPIB Controller. The Instrument is showing as a miscellaneous VISA resource in NI MAX and I am able to send and query simple commands like *IDN? from the VISA Testpanel within NI MAX. However If I try to use the sample VIs from plug and play drivers aquired over the NI Instrument Driver Network I am unable to find the B1500A as an instrument from the VISA resource dropdown menu. If I manually fill in the GPIB adress of the instrument shown in NI MAX (GPIB1::17::INSTR) I get a VISA resource error. Aditionally it has to be noted that the instrument does not show up under connected instruments in the tools>instrumentation overview in Labview.

 

If I retry the same procedure with an NI USB-GPIB-HS controller which I borrowed from another measurement setup, the analyzer is immediately recognized as connected instrument and also shows up in die VISA resource dropdown. The provided sample VIs also work (as far as I've come to test them).

 

Now I ask myself the question how I can get the same unproblematic connection using an Agilent 82357A or B Controller since I have to return the NI USB-GPIB-HS controller in the medium turn. I am also hesitant to spend money on a new NI controller while having three unused Agilent 82357A and B controllers on hand.

 

What I have done so far to tackle this problem:
- reinstall Keysight IO Suite and all NI Software from scratch (with the Keysight VISA installed as secondary in side by side mode)

- activating Keysight 488.2 support on both the B1500A device machine and the remote PC

- activating NiVisatulip passport in die VISA settings

- trying several alternative GPIB board numbers for the different USB-controllers

 

Software Configuration:

Windows 7 Enterprise 64Bit

LabView 2016 64bit
NI-VISA 16.0

NI-VISA runtime 16.0

NI-488.2 16.0

NI-488.2 runtime 16.0

NI DAQmx 16.0.1

NI-Serial Runtime 15.0

Keysight IO Suite 18.0

 

If I omitted any required information I am more than happy to provide it on request.

 

I somehow reached the end of my knowledge and am happy for any suggestion on how to make the connection work using one of the Agilent Controllers (if possible at all).

 

Thanks in advance,

TomTom

 