Can't get the IVI driver to work with Agilent 33522B function generator.(VISA funktions work fine)
I have installed the IVI-C driver. (http://sine.ni.com/apps/utf8/niid_web_display.download_page?p_id_guid=A1C02CABC2854F97E0440021287E6A9E)
It adds an "Instrument Driver software module" ag3352x. (IviDriver 1.1 IVI-C and IviFgen 5.0 IVI-C)
Added a "Logical Name" and selected the "Driver session" (ag3352x) and selected the Hardware resource (GPIB0::11:INSTR)
Tested with the "Simulate With" set to Specific driver works but when i set it to Don't Simulate i get "0xBFF0015 Timeout expired" in LV and TestStand.
The 33522B instrument goes into remote mode and show 2 communication Error: -113 Undefined header, -420 Query UNTERMINATED.
Is this a bug in the IVI driver? or have i missed something.
/J