IVI agilent Funcgen 33522B

$
0
0

 

Can't get the IVI driver to work with Agilent 33522B function generator.(VISA funktions work fine)

 

I have installed the IVI-C driver.  (http://sine.ni.com/apps/utf8/niid_web_display.download_page?p_id_guid=A1C02CABC2854F97E0440021287E6A9E)

It adds an "Instrument Driver software module" ag3352x. (IviDriver 1.1 IVI-C and IviFgen 5.0 IVI-C)

 

Added a "Logical Name"  and selected the "Driver session" (ag3352x) and selected the Hardware resource (GPIB0::11:INSTR)

Tested with the "Simulate With" set to Specific driver works but when i set it to Don't Simulate i get "0xBFF0015 Timeout expired" in LV and TestStand.

The 33522B instrument goes into remote mode and show 2 communication Error: -113 Undefined header, -420 Query UNTERMINATED.

Is this a bug in the IVI driver? or have i missed something.

 

/J

 

 