Hello.
I have a problem and i don't know if its Visa or Hardware error: Whenever i perform a remote sweep on the SourceMeter keithley 2410, adquire more than 292 measurements stored in the trace buffer using the command :TRACATA?, and send another command to the instrument after it, i receive the error -410 Querry interrupted. The funny part is that i actually obtain the data, i can see the huge string, but, again, after i got the data in my hands and send another command, i receive the -410 error, the error dissapears if i turn the instrument off or i set it locally.
It doesn't happen the same when i adquire less than 292 measurements, i send commands normally and i can even query the measurements again and again and the communication doesn't break.
To clarify things even further, i performed a manual sweep with more than 292 measurements, then queried the data using MAX, and then sent another command using MAX, i still receive the error, but, as before, with less than 292 measurements, it performs ok.
I think its very weird, its as if with more than 292 measurements the instruments keeps sending data inifinitely, or Visa asking for data endlessly. And its weirder that it only happens on GPIB communication, using Serial communication it behaves ok.
PD: I ask here and not on the keithley forums because, well, its kind of dead.